The other $5 digital multimeter that I picked up at the Shelby Hamfest was a Fluke 8600A DMM. This one was quite a bit heavier than the 8000A.

Fluke 8600A digital multimeter
Fluke 8600A digital multimeter

Like the 8000A, there’s a label on the bottom that lists the specs and configuration. This unit came with the rechargeable battery option, which would explain the extra heft.

Fluke 8600A specification label
Fluke 8600A specification label

Unfortunately, this one didn’t do anything when the power button was pressed. Getting the cover off the DMM is a simple matter of removing one screw just above the power connector (normally covered by a “Calibration void if removed” type sticker).

The insides are a considerably more complex than the 8000A. A large chunk of the space toward the transformer and power plug end of the board are taken up by the rechargeable batteries. This DMM appears to have had some modifications or repairs made to it.

Inside the Fluke 8600A digital multimeter
Inside the Fluke 8600A digital multimeter

A cursory inspection shows one significant issue: leaked batteries.

Leaky NiCad batteries
Leaky NiCad batteries

At some point, one of the D-cell NiCad batteries was replaced with an AA sized NiCad battery.

The manual I found (dated 1981) says the batteries can be replaced, but the ones in this unit have been tack soldered to the terminals. Perhaps the battery options were changed in later models of the meter.

Another IC (possibly a ROM) was also added to the meter and connected to the board with some ribbon cable. The ribbon cable is soldered to a DM47S188AN (256 bit PROM) and the mystery IC is just attached to the RF shield of one of the daughter boards with some electrical tape. Maybe a repair job, or a modification?

There are four daughter boards in the case that attach to the main board via pin and socket connectors. One board is marked as a battery power supply, which I’d guess takes care of charging the NiCad batteries.

Battery power supply board
Battery power supply board

Board #2 is marked “OHMS CONV ASSY”

Board #3 was marked “INPUT DIV”

Input divider board

Board #4 was a fairly large one with a large shield on it. No clear silkscreen markings labeling the board like the others but according to the manual, it’s the AC converter board. it also had a holder for a spare fuse. Very thoughtful.

AC converter board

Other than the batteries, I don’t see any obvious problems with any of the other components in the meter. I’m thinking maybe if pull out the leaked batteries, that might get the meter running but I’ll need to study the schematics to see how the batteries are connected to everything to figure out if that will work. This thread on the EEVBlog forum has some potentially useful information.

I’ll put this with the rest of the project items in the stack.

Hamfest acquisition: Fluke 8000A DMM

Two of my Shelby Hamfest acquisitions were a couple of Fluke digital multimeters (DMMs) for $5 each. The seller had no idea if they were working or not, but for $10 I figured they would either be handy workbench instruments if they worked, or fun projects if they didn’t. A brief Google search brought me to a nice tear down/repair attempt blog post.

The Fluke 8000A is a pretty nice looking bench DMM with push buttons for function and range selection, and a handle that also doubles as a stand.

Fluke 8000A digital multimeter
Fluke 8000A digital multimeter

A printed label on the bottom provides specifications and indicates what optional features the meter has. Apparently a rechargeable battery was an available option which would have made it handy for field use. This one didn’t come with a battery.

Fluke 8000A specifications label
Fluke 8000A specifications label

A label on the top of the unit says this unit was last calibrated in September 1994, 27 years ago. Someone apparently decided at some point the meter wasn’t working properly anymore and wrote “BAD” on the sticker.

Calibration sticker dated September 1994 on the Fluke 8000A
Calibration sticker dated September 1994 on the Fluke 8000A

Much to my surprise, the meter turned on when I plugged it in. Testing with some precision resistors I have in my collection gave me some pretty good results.

Measuring the DC voltages of some batteries gave me results that compared pretty well with one of my handheld meters. I haven’t tried measuring anything else yet, but it seems like for $5, I’ve got a pretty decently functioning bench meter (at least for most of the things I’d need to do anyway).